Intended Use
Functions
Contraindications
Product Composition
Product Specification and Performance Characteristics
Product Structure
Product Safety Class
Applied Mobile Phones (Network, Data Couplings)
Cautions for Battery Charging
Power ON & OFF
APP Download
Device Connection
Cautions for Using the App
App Troubleshooting
Instruction for Blood Pressure Measurement
Blood Presure Troubleshooting
Calibration of Blood Pressure by Using Static Pressure
Calibration Process
Instruction for Body Temperature Measurement
Body Temperature Troubleshooting
General Warning
Instruction for Blood Oxygen Measurement
Blood Oxygen Troubleshooting
Blood Oxygen Cautions
Instruction for ECG Measurement
ECG Troubleshooting
ECG Cautions
Maintenance
Symbol Definition
EMC Declaration
Thank you for using this Health Monitor. Please read this User Manual carefully to use the medical device safely and correctly.
Intended Use
The health monitor from Qlupod is designed for measuring and transmitting* body temperature, blood pressure, pulse rate, blood oxygen, and ECG.
Functions
The Health Monitor is a wireless device that uses Bluetooth for signal transmission and needs to be connected to a mobile phone application. With the help of infrared and pressure sensors, body temperature, blood pressure, and pulse rate are measured and transmitted via Bluetooth to the app on the phone. The user can view the temperature, blood pressure, and pulse rate in the mobile app.
Contraindications
Product Composition
The Health Monitor consists of the device, cuff and mobile-phone application program. The device is mainly composed of PCB, Plastics, Rechargeable Lithium-Ion Battery, Pump, etc.
Product specification and performance characteristics
The Health Monitor consists of the device, cuff and mobile-phone application program. The device is mainly composed of PCB, Plastics,Rechargeable Lithium-Ion Battery, Pump, etc.
Product Name | Health Monitor | Product Model | HC-03 |
Software | Health Monitor | SW-Version | V0.0.10 |
Battery Nominal Voltage | 3.N | Input Voltage | 5V ± 0.25V |
Battery Rated Capacity | 400mAh | Size | 70mm x 70 mm x 18.4mm |
Working Frequency | 2.4000-2.4835GHz | Net Weight | About 70g |
Communication Protocol | Bluetooth 4.0 | Resting pressure rated range | 0-300mmHg |
Validity period of product | 3 Years | Validity period of cuff | 2 Years |
Lifetime of battery | Charge-discharge cycles 300 times | ||
Bood Pressure measurement | Bood Pressure measurement | Body Temperature measurement | Body Temperature measurement |
Measurement body part | Upper arm | Measurement body part | Forehead |
Circumference | 22-35cm | Measurement scop | 28-42 C |
Measurement scope of systolic pressure | 60-230nvnHg | Measuring error | i0.2C, for 35-42 C i0.4C, for <35 C |
Measurement scope of diastolic pressure | 40-130mmHg | / | / |
Pressure error | ±3mmHg | / | / |
Measurement scope of pulse rate | 40-180bPm | / | / |
Pulse rate error | ±5% | / | / |
Temperature | 5 C-40 C | Temperature | -25 C - +70 C |
Humidity | 15%-93% | Humidity | 93% |
Atmospheric Pressure | 70-106 kPa | Atmospheric Pressure | 50-106kPa |
Product Name | Health Monitor | Product Model | HC-03 |
Software | Health Monitor | SW-Version | V0.0.10 |
Battery Nominal Voltage | 3.N | Input Voltage | 5V ± 0.25V |
Battery Rated Capacity | 400mAh | Size | 70mm x 70 mm x 18.4mm |
Working Frequency | 2.4000-2.4835GHz | Net Weight | About 70g |
Communication Protocol | Bluetooth 4.0 | Resting pressure rated range | 0-300mmHg |
Validity period of product | 3 Years | Validity period of cuff | 2 Years |
Lifetime of battery | Charge-discharge cycles 300 times | ||
Bood Pressure measurement | Bood Pressure measurement | Body Temperature measurement | Body Temperature measurement |
Measurement body part | Upper arm | Measurement body part | Forehead |
Circumference | 22-35cm | Measurement scop | 28-42 C |
Measurement scope of systolic pressure | 60-230nvnHg | Measuring error | i0.2C, for 35-42 C i0.4C, for <35 C |
Measurement scope of diastolic pressure | 40-130mmHg | / | / |
Pressure error | ±3mmHg | / | / |
Measurement scope of pulse rate | 40-180bPm | / | / |
Pulse rate error | ±5% | / | / |
Temperature | 5 C-40 C | Temperature | -25 C - +70 C |
Humidity | 15%-93% | Humidity | 93% |
Atmospheric Pressure | 70-106 kPa | Atmospheric Pressure | 50-106kPa |
Product Structure
Product Safety Class
Protection against electric shock | Class II |
Applied parts | BF |
Protection against harmful ingress of water or particulate matter | IP22 |
Mode of operation | Not continuous operation |
Used in an Oxygen rich environment | Not intended for use in Oxygen rich environment |
Applied mobile phones (Network, Data Couplings)
The Health Monitor should be used together with the application program on mobile phones. The mobile phone should satisfy the following two requirements.
Bluetooth 4.0
Android 4.3 or higher version, iOS 7.0 or higher version.
Package List Accessories
Health Monitor x1, Cuff x1 and User Manual x1
List Battery Charging
Cuff (XD-01) x1 and USB line x1
The color of battery icon on APP means electric quantity status.
Cautions for battery charging
Power ON and OFF
Power ON: press power key 2 seconds. The device shakes slightly. Power on is successful with blue power light. Power OFF: press power key 2 seconds. The device shakes slightly. Power off is successful with power light off.
APP Download
Go to the respective App stores and search for Qlupod.
Device Connection
Press the power key and hold 2 seconds to power on. Open the APP. The device will connect to the APP automatically with the electrical quantity icon appearing on the APP. If the electrical quantity icon is not shown on the APP, click the Bluetooth icon to set up the connection manually.
Cautions for using the APP
Troubleshooting
Malfunctions | Reason | Solution |
---|---|---|
Failure of downloading the APP | Mobile phones do not satisfy the requirements | Check the version of Bluetooth and system of mobile phone |
Failure of automatic connection | Wrong operation sequence | Power on the device first, then log into the APP or connect manually. |
Failure of connection or finding the device | Weak sensitivity of Bluetooth of mobile phone | Power on the device first, then log into the APP or connect manually. |
Instruction for blood pressure measurement
Blood pressure measurement will be influenced by body posture, wearing method of cuff, physical condition, and surrounding environment. The position of the cuff and heart should be kept at the same horizontal level during measurement.
Do not evaluate the result only by your own experience. Please consult the doctor.
Troubleshooting
Malfunctions | Reason | Solution |
---|---|---|
Low reading | Wrong wrapping of cuff: moving, speaking, or wrong sitting posture during measurement. | Wrap the cuff correctly. Adjust the sitting position keep calm during measurement. Please refer to user manual for operation details. |
Failure of inflating | Leakage of cuff or wrong position of device. | Re-position the device. Contact the manufacturer to replace the cuff if the cuff is broken. |
Failure of measurement | Wrong position of device, or inadequate battery charge. | Check the position of the device by referring to the user manual. Check the power quantity. Ensure the battery is sufficiently charged. Please refer to the user manual for operation details. |
Calibration of blood pressure by using static pressure
HC-03 is based on an oscillometer, which means the calculated blood pressure is only accurate if the static pressure is accurate. Therefore, to calibrate blood pressure, static pressure should also be calibrated. FLUCK BP Pump 2 Blood Pressure Simulator is used to calibrate static blood pressure.
Calibration process:
General Warning: You do not need to calibrate the QluPod yourself. Calibration is performed before delivery and is not required again within the validity period.
Instruction for body temperature measurement
Power on the device and open the APP. The device can connect to the APP automatically or manually with the appearance of an electrical quantity icon on the APP.
Keep the forehead dry and clean. Take the device and align the infrared detection head to the center of the forehead. Keep the device 1-2 cm away from the skin. Select "Temperature" in the App. Click "Start" for the measurement. The system will stop automatically after the measurement. The measured value will be shown.
Troubleshooting
Malfunctions | Reason | Solution |
---|---|---|
Big measurement error | Sweat, water, oil etc. on forehead. | Clean the forehead before measurement. |
Big measurement error | Leakage of cuff or wrong position of device. | Keep 1-2 cm distance between detector head and forehead. |
General Warning
Instruction for blood oxygen measurement
Troubleshooting
Malfunctions | Reason | Solution |
---|---|---|
Data changed a lot in a short period | Measured way or body gesture was wrong. | Re-measure according to the instruction. |
Unable to get measured results | Finger overexerted. | Lightly press the middle finger on the device, do not overexert. |
Unable to get measured results | Finger Is too thin. | Use another finger to measure, ensure the finger covering the blood-oxygen light during the process. |
Cautions
Please relax for at least five minutes before measuring. Do not eat, drink alcohol or coffee, exercise, or bathe up to 30 minutes before the measurement.
Please relax as much as possible and do not talk during the measurement procedure.
The health monitor should be placed horizontally during the measurement.
It is recommended to measure with the middle finger, keeping the nails raised. Clean the fingers before the measurement.
DO NOT move the finger during the measurement.
There are some differences in the blood pressure on different fingers, blood circulation, physiological features, hence try to use the same finger to take measurements.
The paralysis state caused by the continuous blood-pressure measurements can influence the blood circulation, which may affect the oximetry values, so do not measure the blood oxygen after finishing the blood pressure measurement.
DO NOT use nail polish on the measuring finger.
DO NOT measure in strong light.
DO NOT measure after exercising.
The measured blood-oxygen value may be incorrect if the tester’s finger skin is too thick or has pigment deposition.
If foreign objects are present between the probe and the parts to be measured, they may
affect the measurement result.
Cold fingertip skin or poor peripheral circulation caused by low temperature treatment or prolonged exposure to low temperature environment can result in insufficient pulse signal, too low measured value or unable to measure.
Instruction for ECG measurement
Troubleshooting
Malfunctions | Reason | Solution |
---|---|---|
ECG Diagram Reverse | The position of 2 hands is reverse. | Please place the monitor correctly in both hands, according to the instructions. |
Cautions
Warning – EMC (Electromagnetic Compatibility)
The device should not be used next to or stacked with other equipment. If it is necessary to use it in such a configuration, the device should be monitored to ensure it operates normally.
Cautions
Maintenance
Do not do service and maintenance while the device is in use.
Please do maintenance using the following methods.
Please store the device using the following methods:
Keep the device away from high temperature, moist, direct sunshine, dust, salty air.
Keep the device away from areas where the working position is unstable or where the device could wobble or collide. Keep the device away from places with
chemicals or corrosive gases. Do not drop the device. Do not fold the cuff too tightly.
Repairs
Do not disassemble an repair by yourself. Please dial the customer service hotline +41 (0) 71 510 05 45 for consultation due to product quality issues or any doubts for measurement results. The cuff is specified for the device. Please contact the manufacturer for the repair of an ar purchase. Do not change it by yourself. Any repairs can only be done by authorized people.
Cleaning and disinfection
Clean the device using a soft, dry cloth. If the device is very dirty, you can use water or a neutral cleaning agent but test it on a small area before cleaning. If necessary, you can use degreasing cotton with ethanol for disinfection.
Environmental protection
When disposing of the device, cuff, or waste battery, please follow local regulations. Please do not discard it casually.
After-sales Guarantee
The guarantee does not apply to the conditions below:
Symbol Definition
EMC Declaration
Guidance and manufacturer’s declaration- electromagnetic emission – for all EQUIPMENT AND SYSTEMS.
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS.
Guidance and manufacturer’s declaration – electromagnetic immunity: The HC-03 is intended for use in the electromagnetic environment specified below. The customer or the user of the HC–03 should assure that it is used in such an environment.
Immunity test | EN 60601 test | Compliance | Electromagnetic Environnement |
---|---|---|---|
Electrostatic Discharge (ESD) IEC 61000-4-2 | ± 8 kV contact ± 15 kV air | ± 8 kV contact ± 15 kV air | The floor should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. |
Electrostatic transient burst / IEC 61000-4-4 | ± 2 kV for power supply lines ± 1 kV for llpuVoutput lines | ± 2 kV for power supply lines ± 1 kV for llpuVoutput lines | Main power quality should be that of a typical commercial or hospital environment. |
Surge IEC 61000-4-5 | ± 1 kV differential mode ± 2 kV c common mode | ± 1 kV differential mode ± 2 kV c common mode | Main power quality should be that of a typical commercial or hospital environment. |
Voltage dips, short interruptions, and voltage variations on power supply input lines IEC 61000-4-11 | <5 % UT (>95 % dip ll UT) for 0.5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dp in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec | <5 % UT (>95 % dip ll UT) for 0.5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dp in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec | Main power quality should be that of a typical commercial or hospital environment. If the user of the HC-03 requires continued operation during power mains interruptions, it is recommended that the HC-03 be powered from an uninterruptible power supply or a battery. |
Power frequency magneticc field (50/60 Hz) IEC 61000-4-6 | 400 A/m | 400 A/m | Power frequency magnetic fields should be at levels typical characteristic location of in a typical commercial or hospital environment. |
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration -electromagnetic immunity- for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING.
Guidance and manufacturer’s declaration – electromagnetic immunity.
The HC-03 is intended for use in the electromagnetic environment specified below. The customer or the user of the HC–03 should assure that it is used in such an environment.
NOTE 1 At 80 MHz and 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetics are affected by absorption and reflection from structures, objects, and people.
a.) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HC-03 is used exceeds the applicable RF compliance level above, the HC-03 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the HC-03.
b.) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [Vl] v/m.
Recommended separation distances between portable and mobile RF communications EQUIPMENT or SYSTEM for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING.
Recommended separation distances between portable and mobile RF Communications equipment and the HC-03.
The HC-03 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HC-03 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HC-03 as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above the recommended separation distanced in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P Is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.